To Submit an order:

Select the ‘Ready to Inject’ Plate type on the order form. From here, you will be allowed to attach a plate map containing your sample information. In the notes section on the order form, please include the following:

  • Which dye set you are using?
  • Has ladder been added, or will we need to add ladder/HiDi?
  • Which run conditions are needed? (From Run Modules table below)

Dye sets we have available:

Our ABI 3730xl machines are calibrated for the DS-31 & DS-33 dye sets.

GeneScan 500LIZ is the only dye set we have stocked on hand to add to samples.

Run Modules:

GS-G5-ShortGS-G5-MedGS-D-ShortGS-D-Med
Dye SetG5G5Any4Dye (LIZ500)Any4Dye (LIZ500)
Oven Temp63636363
Buffer Temp35353535
PreRun_Voltage15.015.015.015.0
PreRun_Time180180180180
Injection_Voltage1.61.61.61.6
Injection_Time15151515
First_Readout_Time200200200200
Second_Readout_Time200200200200
Run_Voltage15.015.015.015.0
Voltage_NumberofSteps10101010
Voltage_Step_Interval20202020
Voltage_Tolerance0.60.60.60.6
Current_Stability30.030.030.030.0
Ramp_Delay1111
Data_Delay500500500500
Run_Time1600200016002000